Measurement
Before lowering the sensor probe into the sample, do not connect the hook on the unit to
a human body.
The correct values are not displayed if the sensor is not mounted when the measurement
display is activated.
Perform DO measurement with no air bubbles in the internal solution.
Do not reuse a membrane cap of DO sensor.
Use the spanner for DO sensor provided to attach or remove the DO sensor.
Avoid both U-53 and U-53G turbidity measurement in air, since the rubber wiper will
quickly become damaged.
Attach sensor guard to sensor probe in the measurement. When sensor probe is used in
calibration cups (black and transparent) or flow cell, the sensor guard can be taken off.
Calibration
During atmosphere calibration for the DO electrode with DO salinity compensation set to
automatic, values are compensated based on electrical conductivity, but calibration is
performed normally.
Location of use and storage
Storage temperature:
10
C to 60
C
Relative humidity: Under 80% and free from condensation
Store the meter in locations void of dust, strong vibrations, direct sunlight, corrosive gases,
near air conditioners or windy areas.
Disposal of the product
When disposing of the product, follow the related laws and/or regulations of your country.