4 Maintenance
90
5.
Attach a membrane cap to DO sensor
6.
Check for membrane surface
Check air bubbles in a membrane cap.
NG
→
Replace a membrane cap again.
Check that span calibration can be performed.
Note
If the membrane cap is not attached correctly, sensitivity may be lost or response speed may
decrease.
with holding a membrane cap tight.
Twist a DO sensor
Good: Limited air bubbles
NG: Air bubbles of more than 5 mm in diameter