4 Maintenance
91
4.7
Troubleshooting
Note
If the sensor probe is removed while the control unit is indicating an error, errors cannot be canceled
by using the ESC key. Either reconnect the sensor probe or restart the control unit.
4.7.1
Error displays
Error
Cause
Solution
Probe ADC error
Internal IC failure
Contact your nearest sales outlet to have the
sensor probe repaired.
Probe EEPROM
error/Factory
Internal IC failure
Turn the power OFF, then restart the system. If
the error persists, initialize the system from the
"System" menu. If the error still persists, contact
your nearest sales outlet to have the sensor
probe repaired.
Probe EEPROM
error/User
Internal IC failure
Turn the power OFF, then restart the system. If
the error persists, initialize the system from the
"System" menu. If the error still persists, contact
your nearest sales outlet to have the sensor
probe repaired.
Turbidity sensor light
source error
Turbidity sensor light source failure
Turn the power OFF, wipe off any water droplets
on the probe, then remove the turbidity sensor.
Check there are no water droplets around the
turbidity sensor connector, then mount the sensor
again. If the error persists, replace the turbidity
sensor.
Turbidity sensor wiper
motor error
The turbidity sensor wiper is not operating.
Press the ESC key. Check there are no obstacles
near the wiper, then perform the measurement
again. If the error persists, the motor will need to
be replaced. Contact your nearest sales outlet to
have the sensor probe repaired.
Probe capacitor error
Low battery voltage or internal IC failure
Turn the power OFF. Replace the display's
batteries. If the error persists, contact your
nearest sales outlet to have the sensor probe
repaired.
Probe EEPROM error Internal IC failure
Press the ESC key, then redo the operation. If the
error persists, turn the power OFF, then restart
the system (the current data will not be saved). If
the error still persists, contact your nearest sales
outlet to have the display repaired.
Probe board error
Probe board failure
Turn the power OFF. Contact your nearest sales
outlet to have the sensor probe repaired.