3 Basic Operation
14
3.1.4
Connecting the control unit and sensor probe
Note
Connect the control unit with its power OFF.
1.
Align the red mark on the connector, and press the connector in until you hear it
click.
2.
Connect the cable’s hook to the display.
3.1.5
Conditioning
Carry out the steps below when using the unit for the first time or when the system has not
been used for 3 months or longer.
1.
Fill the transparent calibration cup to the line with pH 4 standard solution.
The transparent calibration cup has With TURB Measurement and Without TURB
Measurement gauge lines.
2.
Insert the sensor probe in the transparent calibration cup.
Zoom
Red mark
Hook
Transparent calibration cup
Sensor probe
Without TURB Measurement gauge line
With TURB Measurement gauge line