3 Basic Operation
15
Note
Check that all sensors are attached.
3.
Press the control unit's POWER key for about 1 second to turn the power ON.
Leave the unit for at least 30 minutes to condition the sensors.
Note
The operation keys are designed to operate using the pad of a finger, sharp objects can tear the
control unit cover damaging the operation keys.
Tip
The procedure for immersing the sensor probe in the pH standard solution is the same as that
described in “ 3.3.1 Auto calibration ” (page 41).
Auto calibration can be performed using the same pH 4 standard solution that was used in the
conditioning procedure.
Immersing the sensor in the standard solution is generally required for sensor conditioning, but a
voltage supply is required for DO sensor conditioning. Turning ON the power of the control unit is
necessary during sensor conditioning.
DO value will be higher than usual for a while after turning power ON.
See “ 6.4.2 Conditioning of DO sensor ” (page 105).
3.1.6
GPS (U-52G, U-53G, U-54G)
The GPS position measurement precision is proportional to the GPS position measurement
time. When the position measurement precision increases, the position measurement time
also increases. See "
GPS locating accuracy" (page 17) for how to set the position
measurement precision. See "
GPS locate" (page 15) below for how to check acquired
GPS data.
GPS locate
1.
Press the right (
) key to switch the display to the "SETTINGS" screen.
2.
Press the down (
) key to move the cursor to "System", then press the ENTER key.
3.
Press the down (
) key to move the cursor to "GPS locate", then press the ENTER
key.
4.
The message "Press ENT key to start position measurement." appears. Press the
ENTER key.