3 Basic Operation
30
Temperature compensation for conductivity (COND)
Sample conductivity (COND) varies with temperature, and this control unit uses a temperature
compensation coefficient to automatically standardize the conductivity (COND) at 25
C. The
initial setting coefficient is 2%/K, which is the generally used.
1.
Press the control unit’s POWER key for about 1 second to turn the power ON.
The "MEASUREMENT" screen appears after about 10 seconds.
Note
The operation keys are designed to operate using the pad of a finger, sharp objects can tear the
control unit cover damaging the operation keys.
2.
Press the right (
) key to switch the display to the "SETTINGS" screen.
3.
Press the down (
) key to move the cursor to "Compensation", then press the
ENTER key.
4.
Press the down (
) key to move the cursor to "Cond Temp", then press the ENTER
key to toggle the setting between "OFF" and "Input mode".
Default: 2.00%/K