4 Maintenance
89
4.6
Replacing the membrane cap
Replacement procedure
1.
Prepare the DO sensor.
Take a DO sensor out of pack (newly purchasing).
Remove a DO sensor from the sensor probe (after use).
Twist a membrane cap from DO sensor.
Wash the gold electrode and silver electrode parts with water.
2.
Replace the silicone gasket with a new one.
3.
Pour internal solution (#306) into a membrane cap with a dropper.
Check air bubbles in a membrane cap.
4.
Set up a membrane cap on an adjustable mounting.
Newly purchasing
Undo a DO sensor from the sensor probe
Silicon gasket
Pour internal solution till upper line
Pick a Cap up and drop it down, if there is air bubbles in internal solution of it.
Membrane cap
Adjustable mounting
(Spread a soft cloth under an adjustable mounting)