3 Basic Operation
26
3.2.5
Compensation
Note
Compensation settings can only be made when the sensor probe is connected.
U-50 series have following functions of compensation.
Salinity compensation and atmospheric pressure compensation for dissolved oxygen
(DO)
Temperature compensation for conductivity (COND)
Setting total dissolved solid (TDS) coefficient for TDS
Salinity compensation (DO)
The dissolved oxygen (DO) value is presented higher than actual value if salinity
compensation is not added, because the increase of salinity gives higher DO value. To obtain
correct value salinity compensation is needed. The following modes are available for
calculation of salinity compensation.
1.
Press the control unit’s POWER key for about 1 second to turn the power ON.
The "MEASUREMENT" screen appears after about 10 seconds.
Note
The operation keys are designed to operate using the pad of a finger, sharp objects can tear the
control unit cover damaging the operation keys.
2.
Press the right (
) key to switch the display to the "SETTINGS" screen.
3.
Press the down (
) key to move the cursor to "Compensation", then press the
ENTER key.
AUTO:
Salinity compensation is performed automatically with salinity converted from
conductivity.
Value input: Press the up (
) and down (
) keys to enter a setting value when the salinity
is known.