3 Basic Operation
43
6.
Immerse the sensor probe in the transparent calibration cup.
Check that the pH sensor, ORP sensor, reference electrode, COND sensor, TURB
sensor and temperature sensor are submerged in the pH 4 standard solution and
check that there are no air bubbles on the sensor.
7.
With the sensor probe still in the transparent calibration cup, place the transparent
calibration cup into the black calibration cup.
Transparent calibration cup
Sensor probe
Black calibration cup
Transparent calibration cup
Sensor probe