3 Basic Operation
12
Attaching the ORP sensor
1.
Remove the sensor guard.
2.
Remove the sensor plug.
3.
Coat the ORP sensor O-ring with a thin layer of grease (part No. 3014017718).
Note
Be sure no grease from the O-ring gets on the sensor connector. If the sensor connector gets grease
on it, wipe it off with a soft cloth soaked in alcohol.
4.
Make sure there is no moisture on the sensor probe’s sensor connector (marked
"ORP").
5.
Fasten the ORP sensor securely by hand.
6.
Clean the sensor with an alcohol-soaked cloth.
Attaching the reference electrode
1.
Remove the sensor guard.
2.
Remove the sensor plug.
3.
Coat the reference electrode O-ring with a thin layer of grease (part No.
3014017718).
Note
Be sure no grease from the O-ring gets on the sensor connector. If the sensor connector gets grease
on it, wipe it off with a soft cloth soaked in alcohol.
4.
Make sure there is no moisture on the sensor probe’s sensor connector (marked
"REF").
5.
Fasten the reference electrode securely by hand.
6.
Remove the tape from the liquid junction part of the reference electrode.
ORP sensor
Reference electrode
Tape
Содержание U-51