3 Basic Operation
42
2.
Remove the transparent calibration cup.
3.
Fill the transparent calibration cup to the line with pH 4 standard solution.
The transparent calibration cup has With TURB Measurement and Without TURB
Measurement gauge lines.
4.
Press the control unit’s CAL key to set the calibration mode.
5.
Press the down (
) key to move the cursor to "Auto Calibration", then press the
ENTER key.
Transparent calibration cup
Without TURB Measurement gauge line
With TURB Measurement gauge line