3 Basic Operation
49
6.
Press the up (
) and down (
) keys to set the mV value of the ORP standard
solution containing the submerged sensor probe at the measurement temperature.
Table 3 Indicated value of ORP standard solution at various temperatures (mV)
7.
Check that "Measurement value" has stabilized, then press the ENTER key to start
calibration.
8.
Calibration is finished when the message "Cal complete. ENT to manual cal menu."
appears. Press the ENTER key to return to the calibration parameter selection
screen.
Temperature
160-22
16051
5
+274
+112
10
+271
+107
15
+267
+101
20
+263
+95
25
+258
+89
30
+254
+83
35
+249
+76
40
+244
+69