5.1.3. Supply module label
To replace the 2 fuses located on the left of the power supply connector on the back of the analyser,
carry out the following procedure:
◆
Do not remove the instrument protection cover.
◆
Switch the instrument off.
◆
Disconnect the main power cable from the back power cord receptacle on the instrument.
◆
Pull the little flap marked "250V fuse".
◆
Remove the fuses from their holding receptacle
◆
Check for the correct ohm value on each fuse.
◆
Use only "Slow-blow" internal fuses.
◆
Use only fuses having the following characteristics:
for 100/240Vac supply: F1-F2 T2AL250V
5.1.4. Diluent and waste connection label
Waste
: connect the waste output line to the (waste position) fitting. Note the waste label for waste
output "Only".
Diluent
: connect the diluent input line to the (diluent position) fitting. Note the diluent label for
diluent input "Only".
Diag.1-5 Supply module label
1-
ON/OFF switch
2-
Fuses location
3-
Power supply connector
Diag.1-6 Supply module label
1-
Diluent input connector
2-
Waste output connector
3
2
1
2
1
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...