Maintenance &
Troubleshooting
User Manual - RAB237AEN
7–15
4-
The free male connector must be plugged onto the pack upper valve in order to receive the waste
liquids (
).
5-
to run the priming after the pack replacement.
3.5.4. ABX Diluent & waste connections (Bottle version only)
◆
Connect the ABX Diluent container using the diluent straw and a 3x6 cristal tube (1 meter
maximum) on the diluent input (located at the bottom of the instrument rear panel) and place
the waste container below the instrument level (under the bench).
Diag.7-22 Waste connector
It is recommended not to remove the pack several times before the reagents
are totally used to avoid leak problems.
Diag.7-23 Rear connections
Always follow the recommended procedures for waste disposal. Never connect
the instrument wastes directly to the laboratory drain pipes.
The ABX Diluent container will be located on the bench at the same level as
the instrument.