ABX Micros
ES60
OT/CT
7–6
User Manual - RAB237AEN
3.2. Cover dismantling
◆
Using the key from the installation kit, turn the locker (
) to open the pneumatic
protection door.
◆
Unscrew the 3 cover fixation screws (
Diag.7-5-B
) and loosen the 2 screws (
) under the
reagent flap.
◆
Disconnect the integrated printer connectors when removing the cover (
◆
Remove the cover: pull it backward and lift it up to the rear of the instrument.
Diag.7-5 Cover dismantling
Diag.7-6 Printer connectors
B
B
A
B
C
C