Specifi
cations
User Manual - RAB237AEN
2–17
4. Reagent
4.1. Reagent specifications
◆
Do not use the instrument operational caps when reagents are left off the instrument.
◆
The HORIBA ABX reagents specified for this instrument have been approved in accordance with
the European Directive 98/79/CE (Annex III) for in-vitro medical devices.
◆
Refer to reagent leaflets for all reagent specifications.
◆
Allow your reagents to return to room temperature before use, to avoid gas bubble emissions.
◆
Never pour the remaining quantity of a reagent previously used into the new reagent replacing it.
This will eliminate any cross contamination to the new reagent.
◆
Never leave a reagent container open during use. Use the appropriate caps provided with the
instrument. Order new reagent caps if lost or misplaced.
◆
Check with your shipping company and verify that your reagent shipment has not encountered
important temperature differences during transportation.
◆
Clean any reagent spillage with water as soon as possible to prevent crystallization of the reagent
and oxidation of the metal parts of the instrument.
◆
Never pour reagents into the laboratory waste water drainage system. Follow local/national
regulations for chemical waste disposal. Apply neutralization procedures when necessary.
◆
It is necessary to flush the reagent lines with distilled water, then dry them when the instrument
is going to be shipped to any location or will be left without operating for an extended period of
time. Contact your local HORIBA ABX Technical Support Representative for more information about
this procedure.
◆
These reagents are used for Invitro diagnostics.
4.2. Reagent consumption
◆
Reagent consumption is given in millimiters (ml). It has been calculated according to a specific
software version.
Tab.2-15 Reagent consumption on ABX Micros
ES60
CT (software version V1.1.X)
HORIBA ABX manufactures reagents, calibrators and quality control bloods
specially designed for use with the ABX Micros
ES60
OT/CT analyser. The use
of products not recommended by HORIBA ABX may give erroneous results or
instrument operation problems. Contact your local HORIBA ABX Technical
Support Representative for all information regarding the recommended
products.
Cycle
Diluent
Lyse
Cleaner
ABX Minoclair
Startup
23.5
0.5
-
-
Shutdown
-
-
14.4
-
Analysis
16.8
0.5
1.0
-
Prime all reagents
26.4
10.3
5.4
-
Automatic cleaning
16.3
0.5
13.8
-
Concentrated cleaning
16.3
0.5
1.9
6.0
Cal photometer
16.3
0.5
1.6
-
Backflush
-
-
-
-
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...