4.2.4. Reagent priming
This function primes all reagents of the pack in pack mode and the selected reagent in the bottle
mode.
In the "
Reag.
" tab:
1-
Press "
Prime
" button (
).
4.2.5. Reagent closing
A reagent has to be closed when a reagent is expired.
Bottle unit
In the "
Reag.
" tab:
1-
Press corresponding radio button ("
LYSE
", "
DILUENT
" or "
CLEANER
") to select the reagent to
close.
2-
Press "
Delete
" button (
).
◆
The following message is displayed: "
Are you sure you want to close the selected reagent?
"
3-
Press "
OK
" button.
Pack unit
In the "
Reag.
" tab:
1-
Press "
Delete
" button (
).
◆
The following message is displayed: "
Are you sure you want to close the selected reagent?
"
2-
Press "
OK
" button.
All the reagents of the pack are closed.
Diag.7-31 Prime button
A priming cycle must be done after each reagent installation (creation or
replacement).
Diag.7-32 Delete button
When a reagent is closed, the following information is displayed in the "
Logs
"
tab in the "
Reagent
" category:
- the date/hour of the reagent closing
- the user who closed the reagent
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...