Intro
duction
User Manual - RAB237AEN
1–5
2. Operational conditions
2.1. Environment
◆
The operation of the ABX Micros
ES60
OT/CT should be restricted to indoor location use only. This
instrument is operational at an altitudes of maximum 3000 meters (9800 feet) in bottle version
and 3000 meters (9800 feet) in pack version.
◆
The ABX Micros
ES60
OT/CT is designed for safety from voltage surges according to INSTALLATION
CATEGORY II and POLLUTION DEGREE 2 (IEC 61010-1 (2002)).
2.2. Location
◆
The ABX Micros
ES60
OT/CT should be placed on a clean and leveled table or workbench.
◆
Please note that the instrument, integrated printer and reagents weigh approximately 20 kilograms (44 lbs).
◆
Avoid exposure to sunlight.
◆
Place your instrument in a well-ventilated area.
◆
Place your instrument where it is not exposed to water or vapor.
◆
Place your instrument where it is free from vibration or shock.
◆
Place your instrument where an independent power receptacle can be used.
◆
Use a receptacle different from the one used by a device that easily generates noise such as a
centrifuge, etc...
◆
Provide a space of at least 20 cm (8 inches) at the back of the instrument for arranging the power
cable and tubings.
2.3. Grounding
Proper grounding is required when installing the system. Check the wall outlet ground (Earth) for
proper grounding to the facilities electrical ground. If you are unsure of the outlet grounding,
contact your facilities engineer to verify the proper outlet ground.
2.4. Humidity/temperature conditions
◆
The ABX Micros
ES60
OT/CT must operate between temperatures of 16°C to 3
0
°C (61°F to
86
°F).
Maximum relative humidity 85% for temperatures up to 3
0
°C (
86
°F) without condensation.
◆
If it is stored at a temperature less than 10°C (50°F), the instrument should stand for 1 hour at
the correct room temperature before use.
2.5. Electromagnetic environment check
◆
The ABX Micros
ES60
OT/CT has been designed to produce less than the accepted level of
electromagnetic interference in order to operate in conformity with its destination, allowing the
correct operation of other instruments also in conformity with their destination.
◆
In the case of suspected electromagnetic noise, check that the instrument has not been placed
in the proximity of electromagnetic fields or short wave emissions, (i. e. Radar, X-rays, Scanners,
Cell phones, etc...)
2.6. Main supply
◆
Grounding is required. Check that the earth wall-plug is correctly connected to the laboratory
grounding system. If there is no such system a ground stake should be used.
◆
Use only main supply cable delivered with the ABX Micros
ES60
OT/CT.
◆
Main supply voltage fluctuations must not /-10% of the nominal voltage.
Voltage input: ..100V to 240V (+/-10%)
Frequency: .......50Hz to 60HZ
Power: ... ..........110VA max.
Please contact your local HORIBA ABX
Technical Support Representative if use
conditions do not comply with the recommended specifications.
The power switch and input voltage supply connection should always be
accessible. When positioning the system for operational use, leave the
required amount of space for easy accessibility to these items.
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...