ABX Micros
ES60
OT/CT
5–24
User Manual - RAB237AEN
3-
At the end of the cycle, a dialog box displays the calibration sample run results. Press "
OK
" button
to validate results or press "
Cancel
" button to refuse results.
4-
Place the cap back onto the vial, gently and thoroughly mix the material for the next calibration
sample run.
5-
Run the remaining calibrator samples (See
Running calibration, page 5-23
Wipe any excess blood from the cap and threads of the calibrator vial with a
lint-free tissue to prevent any dried blood from re-entering into the calibrator
material. Dried blood re-entering into the vial may give error flags and reject
the sample runs.
If a calibration sample run result is refused by the user, the following message
is displayed: "
You refused the result obtained. You can start your analysis
again.
".
Press "
OK
" button. The following message is displayed: "
Last result refused.
Please insert your sample and launch the 1/X analysis of the calibration
procedure.
". Launch the calibration cycle.
Calibration results having error flags such as ($, *, or ! for HGB) will
automatically be rejected! The system will automatically re-set itself to re-
run that sample. If you acquire 3 rejects continuously on the same number
sample, abort calibration and contact your local HORIBA ABX
Technical
Support Representative!
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...