9. Installation Report
Technical Manual P/N 9894.00.G6
9-7
9.4
INITIAL TURN-ON
9.4.1
Preliminary Generator Power-On Test
Pass ( )
Fail
( )
(Section 4.1)
9.4.2
Exposure Test Verification
System Not Calibrated : Exposure Test Verification [Section 5.1.2]
Exposure
Measured values
Tube loading range
Test
50 kV
10 mA
200 ms
______ kV
_______ ms
45.5
≤
kV
≤
54.5
189.8
≤
ms
≤
210.2
Pass ( ) Fail ( )
Pass ( ) Fail ( )
System Calibrated : Exposure Test Verification [Section 5.3.1]
Exposure
Measured values
Tube loading Range
Test
100 kV
200 mA
100 ms
______ kV
______ mA
_______ ms
93
≤
kV
≤
107
189.5
≤
mA
≤
210.5
94.8
≤
ms
≤
105.2
Pass ( ) Fail ( )
Pass ( ) Fail ( )
Pass ( ) Fail ( )
mAs Verification [Section 5.3.2]
2-point exposure
80 kV; 200 mAs
AEC Exposure
80 kV; 200 mA
Displayed mAs : _____ mAs
Displayed Backup mAs: _____ mAs