5. Field Calibration
Technical Manual P/N 9894.00.G6
5-7
Tube Calibration Messages
-
"Press Prep & RAD to begin": The generator is ready to calibrate the selected filament; press the
console exposure buttons (or room footswitch) to start the calibration.
-
"Searching Emission": The generator searches for the proper current value that will emit tube
current.
-
"Acquiring mA Samples": The generator does exposures and reads mA values.
-
"I FIL : X.X Amp ": Display of filament current delivered for the current kV station.
-
"Calib Done ": The generator has completed the calibration for the selected filament
-
"Calib Aborted ": The operator has cancelled the tube calibration (Cancel key).
-
"No emission ": Error message
-
"No mA sample ": Error message
Tube Calibration Viewer
It is possible to see the different tube emission calibration curves resulting from a tube calibration.
The different curves represent the plotting of the tube current (in mA) versus tube filament current
(in Amps) for different kV stations. This is done using the Epsilon PC tools module; refer to Section
7.5.3.3.
Figure 5.4
–
Tube Calibration Viewer.