6. User Acceptance Tests
Initials: _____
Technical Manual P/N 9894.00.G6
6-5
Table 58-
Tube mA Accuracy (Continuous mode)
Requested Conditions
Exposure 1
20% mA
MAX
Compliance Criteria
Lowest kV
____ mA
___ mA ± (5% + 0.5 mA)
____ kV
Select
Measure ____ mA
Test Passed
Test Failed
Requested Conditions
Exposure 2
20% mA
MAX
Compliance Criteria
Highest kV
____ mA
___ mA ± (5% + 0.5 mA)
____ kV
Select
Measure ____ mA
Test Passed
Test Failed
6.2.3
mAs Accuracy
Acceptance Test Logbook
(4/11)
Perform the two mAs exposures (2) using the loading factors of Table 59 and
verify the accuracy of the tube mAs.
Table 59
Passed
Failed
Table 59-
Tube mAs Accuracy (Intermittent mode)
Requested Conditions
Exposure 1
Lowest mAs
Compliance Criteria
Highest kV
0.4 mAs
0,4 mAs ± (5% + 0,2 mAs)
____ kV
Select
Measure ____ mAs
Test Passed
Test Failed
Requested Conditions
Exposure 2
Highest mAs
Compliance Criteria
Lowest kV
____ mAs
___ mAs ± (5% + 0,2 mAs)
40 kV
Select
Measure ____ mAs
Test Passed
Test Failed