EPS 45-80 High-Voltage Generator
7-24
Technical Manual P/N 9894.00.G6
o
Inverters
zone: Gives the status of the 3-Phase Inverter Boards. See also Section 7.5.2.6.
1.
First column (Board)
3-Phase Inverter board number, from position 1 thru position 6, as installed on the
Motherboard.
2.
Second column (Status)
Indicates the status of the Inverter board installed at the given position (Table 83).
3.
Third column (Sel.)
Indicates which Inverter boards are presently selected for the given generator's state
shown in the State zone. Not all boards are used at all times.
4.
Fourth column (Temp.)
Indicates the estimated temperature at the surface of every Inverter boards (34°C in Figure
7.12). The ambient temperature is also given (22°C at the sensor location).
Fix button
: Use this button to reset the Inverter Status Register, for instance when an
Inverter board is replaced. The generator does a re-initialization process to validate the
reset.
Table 83-
Possible 3-Phase Inverter Status in EEPROM Register
Status
Meaning (for the given Inverter Slot position)
OK
A functional Inverter board is installed in this slot position.
FUSE
The Inverter board in this position has a blown fuse but is still usable by the generator.
DUMMY A Dummy board is installed in this inverter slot position and the output kW is
therefore derated.
BAD
The Inverter board in this position is in unusable condition. The output kW is
therefore derated.
STOP
Unusual condition, the generator won't allow exposures. Verify the corresponding
Inverter board inside the generator.
o
I/O
zone: Gives the status of the different programmable inputs and outputs for the given
generator's state shown in the State zone.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...