EPS 45-80 High-Voltage Generator
7-26
Technical Manual P/N 9894.00.G6
Other information or actions can be obtained from the History Log screen:
Set Date and Time
: Use this button to synchronize the generator clock with the PC.
Exposure Counters
: Number of exposures done on both filaments.
Firmware
:
Names, part numbers and revision numbers of installed firmware.
Arc Counters
:
Number of tube arcs (for AG, AK and KG figures).
AG : Anode-to-Ground
arcs
AK : Anode-to-Cathode
arcs
KG : Cathode-to-Ground
arcs.
Inv. Max. Temp.
:
The maximum temperature every 3-Phase Inverter board has reached
during the history log period.
HU Max.
:
Number of times an exposure was not allowed because it would otherwise
have exceeded the Maximum Tube Heat Units.
HU 100%
:
Number of occurrences of Tube Heat Units greater than a hundred
percent.
Housing
:
Number of occurrences of Tube Housing Inhibit conditions.
Over kV
:
Number of occurrences of Over KV fault.
Power Quality
The
Power Quality
figure has an effect over the maximum kW admitted
for an exposure. The power quality percent is measured and refreshed
during idle mode. It varies from about 80% to about 125% and is based on
the following instantaneous four (4) measurements or values:
1. Number of good Inverters.
2. Maximum kW Figure set in the generator configuration thru
the Options setup in Epsilon PC Tools.
3. Instantaneous Line Voltage.
4. Measured Line Impedance.
•
When the power quality figure reaches or exceeds 100%, the generator will be capable of
doing the most demanding exposure based on the maximum power output of the
generator configuration and inverter board status.
•
When the power quality figure falls below 100%, the generator might not be capable of
doing certain demanding exposures
.
This could be the result of a bad Inverter, a bad line
condition, too much impedance on line or other. When the generator inhibits a kW-
demanding exposure, it displays the message “Generator kW Limit” at the operator
console.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...