EPS 45-80 High-Voltage Generator
4-32
Technical Manual P/N 9894.00.G6
III.
I
MAGE
E
NHANCEMENT
(O
PTION
): Select this option to enable
Image Enhancement mode.
Do not adjust the Enhancement factors yet.
This setup will be done later (Section 5.4.3).
•
SPD
(O
PTION
):
If Source-To-Patient Distance tracking is interfaced with the generator,
enable the SPD tracking function.
•
A
NODE
S
PEED
:
Choose the anode speed:
1. RAD Speed;
2. Low Speed; 3. High Speed. For faster transition from Fluoro to RAD,
select RAD speed for the anode rotation.
•
A
LARM
T
IMER
:
Set the Timer for the Fluoro alarm (up to 5 minutes).
•
P
ULSE
R
ATE
: For optimum friendliness in Pulsed Fluoro, enable only those pulse rates that
will be used.
•
M
IN
./M
AX
./S
TARTING K
V:
Use these settings to configure the kV range and starting kV in
Fluoro. The starting kV is used at the beginning of a new patient exam only.
•
M
AXIMUM M
A
S
MALL
F
ILAMENT
:
This selection permits to specify a maximum mA threshold
value for the Fluoro small filament, beyond which Fluoro would switch to the large filament.
This value should be considered together with the adjustment of the Maximum Fluoroscopic
Entrance Exposure Rate (Section 5.4).
•
S
OURCE FOR
ABS
C
ONTROL
:
Identify the source for the ABS Fluoro dose stabilization:
ION # 1
(1)
(Controller board, J13)
ION # 2
(1)
(Controller board, J8)
ION # 3
(1, 3)
(J1 connector)
ION # 8
(1, 5)
(J5 connector)
ION # 9
(1, 5)
(J4 connector)
Remote AEC
(2)
(Controller board, J14 pin 10)
Solid State Chamber # 4
(1, 3)
(J3 connector)
Solid State Chamber # 5
(1, 3)
(J4 connector)
Solid State Chamber # 6
(1, 3)
(J5 connector)
PMT (Fluoro PMT board)
(4)
(J5 connector)
Photodiode (PhD)
(5)
(J6 connector)
Notes:
1. For Pulsed Fluoro only.
2. For
Continuous
Fluoroscopy only.
3. On
Universal
Chambers Interface Board P/N 11002.00
4. Select the source for the PMT gain adjustment:
i. Potentiometers located on Fluoro PMT board
ii. Software (uses an internal D/A to set the gain value)
iii. External (J3B, pin 3 on Fluoro PMT board)
When a PMT device is used for both ABS stabilization and AEC control, do not select the
same PMT gain adjust source for Fluoro and RAD. If High Level Fluoro is done, select two
different sources for adjusting the gain, one for low level, the other for high level.
5.
On Photodiode Ion Chamber Fluoro board P/N 11030.000.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...