Introduction to
Epsilon
Technical Manual
Technical Manual P/N 9894.00.G6
v
B. SAFETY PROCEDURES
Disregard of the procedures given below may cause serious or fatal
bodily injury.
B.1 Electrical Safety
All of the movable assemblies and parts of this equipment should be operated with care and
routinely inspected in accordance with the manufacturer's recommendations contained in the
present manual.
Only properly trained and qualified personnel should be permitted access to any internal parts. Live
electrical terminals are deadly; be sure line disconnect switches are opened and other appropriate
precautions are taken before attaching accessories.
Do not remove the flexible H. T. cables from the X-ray tube housing or Anode or Cathode
assemblies H.T. connector until the electrical mains have been disconnected for
at least
1 minute.
B.2 Radiation Safety
X-rays are dangerous to both operator and others in the vicinity unless established safe exposure
procedures are strictly observed.
The useful and scattered beams can produce serious or fatal bodily injuries to any persons in the
surrounding area if used by an unskilled operator. Adequate precautions must always be taken to
avoid exposure to the useful beam, as well as to leakage radiation from within the source housing or
to scattered radiation resulting from the passage of radiation through matter.
Those authorized to operate, participate in or supervise the operation of the equipment must be
thoroughly familiar and comply completely with the currently established safe exposure factors and
procedures described in publications. Failure to observe these warnings may cause serious or fatal
bodily injuries to the operator or those in the area.
B.3 Class l Equipment
According to the class of equipment, the Epsilon High-Voltage generator must be referenced to
an earth ground by a protective earth conductor. Failure to provide a separate earth ground can
result in electrical shock hazard causing injury or death
.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...