2. Installation
Technical Manual P/N 9894.00.G6
2-7
2.4
POWER LINE CONNECTION
The generator requires a three-phase electric line with a protective earth ground (4 or 5 wires,
maximum wire gauge 0 AWG (50mm)).
Refer to the Power Rating Table (Section 1.2.4, Table 10)
to make sure the power line ratings and electrical safety
requirements are met.
=
>>
1.
Make sure the mains line is turned off
at the room circuit breaker.
2.
Remove the Line Access Panel at the back of the
Epsilon
High-Voltage Generator to feed the
line wires through the cable clamp (Figure 2.3).
3.
Connect the mains line of the disconnect switch to the Line Filter Board:
J2
(
φ
1),
J3
(
φ
2),
J4
(
φ
3)
and
J5
(Neutral)
(i)
lugs
(ii)
(Figure 2.4).
(i)
For
Delta
line configuration, leave the neutral lug unconnected.
(ii)
Lugs maximum wire gauge: 0 AWG (50 mm
2
).
4.
Connect the main ground of the X-ray room to
J1
lug.
Failure to provide a separate earth ground can result
in electrical shock hazard causing injury or death
.
5.
Firmly attach the wires on the Line Filter Board Holder using cable-ties.
6.
Do not turn on the mains power
to the generator at this moment.
Figure 2.4
Line Filter Board
i) Power Line connection
ii) Tube Stator connection
iii)
Tube
Housing
connection
iv)
K1/K2
Relay
connections
v) F3 & F4 fuses are installed with
Auxiliary Power Supplies
option
only.
See Table 10,
In Section 1.2.4.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...