3. Interfacing
Technical Manual P/N 9894.00.G6
3-39
3.8
COMMUNICATION PORTS
3.8.1
Communication with a Computer
Serial communication with a computer can be established using J16 port of Controller board or J6
port of the optional Epsilon Operator Console. Both serial ports can be used by the Epsilon PC
Tools
program to
configure
or
service
the generator. Connect a standard RS-232 cable to link the
generator to the computer (Table 47).
Table 47-
RS-232 Communication Ports
Generator Serial Ports
Computer Serial Port
Pin
Controller J16
Console J6
DB9
1 GND
N.C.
N.U.
2 TX
TX
RX
3 RX
RX
TX
4 DTR
N.C.
N.U.
5 GND
GND
GND
6 GND
N.C.
N.U.
7 RTS
RTS
RTS
8 CTS
CTS
CTS
9 GND
N.C.
N.U.
Baud rate :
57600
WARNING
Do not make an exposure
when a computer or any external equipment is directly
connected to the operator console (
J6)
or generator (
J16)
.
If exposing is necessary, use an optical isolator between the computer and the
console.
Recommended part
: Optical Isolator, RS-232 to RS-232,
model
9SPOP2
, from B&B Electronics
It is not recommended to connect
a computer or any external equipment to the operator
console (J6) or generator (J16) without the use of an opto-isolator, unless that
equipment is battery-powered.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...