EPS 45-80 High-Voltage Generator
7-52
Technical Manual P/N 9894.00.G6
7.6.5
Software Update
7.6.5.1
Firmware & Epsilon PC Tools
Use the following procedure to update the software on an installed Epsilon generator. The Epsilon
software consists of three firmware S19 files (S19 file format: low-level machine code and addresses)
and one Epsilon PC Tools executable file.
°
(1) X-ray P/N 9809.01, (2) CIO P/N 9810.01
and (3) Console P/N 9811.xx firmwares
°
Epsilon PC Tools Program P/N 9893.01
The four items form a group, and an element of one group or revision cannot be mixed with an
element of another group or revision. When updating a firmware, make sure that it is compatible
with the other two firmwares and with the current Epsilon PC Tools revision. If necessary, call
emd
technologies
for additional information, or verify the software compatibility on your customer web
page.
No
tes
Epsilon Loader
This module, when used inappropriately, can lead to serious or fatal impairment
of the generator. Do not use this module unless you are sure to have a backup or
an adequate copy of the generator firmware (Section 8.2.3.5).
Updating the CIO or Console firmware will result in the replacement of important
configuration data! Make sure to have a backup of the generator configuration data before
loading new software.
Firmware
Devices Important Notes
9810.01.S19
CIO
Will reset all the hardware setup, the tube calibration and the Fluoro dose
calibration.
9809.01.S19
X-RAY Configuration data not replaced.
9811.xx.S19
Console Will replace all APR data by the cold load parameters.
If you do not have the firmware and Epsilon PC Tools with you on a CD or on your laptop, you
will have to download them from
emd
technologies
web site using the following procedure:
a.
Go to the emd web site :
www.emd-technologies.com
b.
Enter your Client Name in the Customer Name Zone and click on
OK
.
c.
Enter your password and click on
Submit
.
Note:
Client Names & passwords are case sensitive!
d.
You will get access to the latest documentation and software/firmware available for your application.
e.
Download the “Epsilon PC Tools.exe” and save this file on a specific folder.
f.
Download the Firmware zip file (Firmware.XX.ddmmm.yy.Zip) on the same folder.
XX=04 => RAD-only firmware package; XX=05 => RF firmware package
g.
Extract the zip file into its three (3) firmware constituents: XRAY, CIO and CON.
Unbreakable
Software
Group
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...