3. Interfacing
Technical Manual P/N 9894.00.G6
3-17
Table 40-
Fluoro Switch Interface
Board part number
Gray
= Optional connectors
Signal Name
PHD
11030.000
PMT
9822.00
Description
Fluoro High Level
J2-1
J2-3
Low for Fluoro High level
Fluoro Low Level
J2-2
J2-4
Low for Fluoro Low level
Footswitch common
J2-3
J2-5
Return signal for the Fluoro switch
3.3
EXPOSURE TRIGGERS
The generator provides 2 RAD trigger inputs (J14 & J15, pin 6 in Figure 3.3a) and one trigger input
for Fluoro to synchronize the exposure with the peripheral equipment. The trigger pulse width must
be > 250 µs. The activation voltage depends on V
TH
(Section 3.4.1.2).
To assign the proper trigger input and configure its activation level for a given workstation, use the
Epsilon
PC Tools Program (Section 4.5 for RAD & Section 4.6 for Fluoro). Refer to Figure 3.3b
for different types of trigger activation.
Notes: 1. Utilization of a trigger in RAD is mandatory in serial radiography only.
2. Refer to Section 3.7 for Fluoro trigger.
Figure 3.3a –
RAD Trigger Interfacing.