Technical Manual P/N 9894.00.G6
6-1
6.0
USER ACCEPTANCE TESTS
This section sets forth the acceptance testing criteria that should be used to determine compliance of
the
Epsilon
High-Voltage generator with the general performance requirements of the
manufacturer. It is not intended to provide complete conformance evidence for all applicable
regulatory requirements.
The testing procedure applies to the
final
generator, that is, to the generator that has been assembled,
installed, programmed and calibrated as per the instructions given in this manual, Sections 2 thru 5.
PRE-REQUISITE CHECKUP
•
Make sure the Epsilon generator is installed as per Chapters 2 & 3 of this manual.
•
Make sure the Epsilon generator is setup and calibrated as per Chapters 4 & 5 of this
manual.
•
Make sure the Installation Report is completed as per Chapter 9 of this manual.
TERMINOLOGY AND DEFINITIONS
ABS
: Automatic Brightness Stabilization, in Fluoro mode. Mode of operation in which the
factors of the Fluoroscopic beam is automatically adjusted to obtain optimal image
brightness.
AEC
: Automatic Exposure Control, in RAD mode
In an X-ray Generator, mode of operation in which one or more loading factors are controlled automatically in
order to obtain at a preselected location a desired quantity of radiation
(1)
kV
: Peak voltage applied between the anode and cathode of an x-ray tube, in thousands of
volts.
kW
: Describes the product of both
kV
and
mA
output at the generator or loaded at the
terminals of the X-ray tube.
mA
: Average x-ray tube current, in milliamperes, during the irradiation time. The operator
console can also display the peak mA rather than the average mA, during the optional
Pulsed Fluoro, if the user does want so.
mAs
: X-ray tube current by time product (
mA x Time
), in milliamperes-seconds
SPD
: Source to Patient Distance
Time
: Irradiation time, in milliseconds (ms) or seconds (s). Irradiation time represents the time
interval between the instant that the tube potential has risen for the first time to a value
of 80% and the instant at which it finally drops below the same value.
WKS
: Workstation (or image receptor). Transducer device that converts the X-ray beam energy
across the patient into visible images.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...