7. Troubleshooting & Servicing
Technical Manual P/N 9894.00.G6
7-25
7.5.3.2
History Log
This screen presents different events that occurred during the operation of the generator; the events
being tracked and logged in a FLASH memory and also displayed in service mode (Section 7.5.2.5).
Figure 7.13 below shows that events are encoded using a 2 byte-code (6F, 7B, 21, …etc), are
accompanied of a description (History Deleted; New Configuration Loaded,….), the IGBT status
(5000, 5900, ….) at the time the event was logged (June 28, 2006) and the devices that have captured
the events (XRAY µp, CIO µp or CON µp).
The data column displays internal information only.
Figure 7.13
–
History Log of the Epsilon PC Tools' Service Module.
History data must be fetched from the generator using the
Query
button when the PC is connected
to the generator. The log can be deleted (
Delete
) or it can be exported (
Export
) as a text file for
later reference. When the log is deleted, all events are deleted from the log and the following statistic
figures (displayed in Service mode, see Section 7.5.2.7) are zeroed:
1. Maximum Tube H.U.;
2. Tube H.U. greater than 100%;
3. Tube Housing;
4. Over KV;
5. Maximum Inverter Temperature.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...