EPS 45-80 High-Voltage Generator
4-34
Technical Manual P/N 9894.00.G6
•
A
DDITIONAL
P
ULSED
F
LUORO
S
ETUP
:
If Pulsed Fluoro is fitted on your generator, perform the
following additional setups.
I.
Select the Fluoro mA display mode (Figure 4.18):
1. Average mA; 2. Peak mA.
Peak mA Display
Average mA Display
If "mA peak Fluoro display mode" is selected, peak mA values will be displayed on the
Epsilon
Console together with the
pulse width parameter (ms) during Fluoro.
Figure 4.18
–
Pulsed Fluoro mA Display modes.
II.
Configure the pulse rate operation (Figure 4.17).
1.
Set the pulse rate values using the horizontal selection bars and enable their selections
using the checkbox located at their left. For optimum friendliness, enable only those
pulse rates that will actually be used.
2.
If only one pulse rate is available, you can disable the display of the pulse rate on the
operator console in Fluoro by deselecting the "Display F/R" checkbox.
3.
If Fluoro pulse rate have to be transmitted to or received from the imaging system,
assign the proper bit configuration for the communication (up to 3 bits).
4.
Save the data and send it to the generator (Section 4.2.3).
4.6.2
Fluoro ABS Curves
For the moment, do not adjust the ABS curves. Wait until Section 5.6 to program
the Fluoro kV/mA tracking curves.
4.6.3
Multi-Zoom Fluoro Dose Stabilization
For the moment, do not adjust the ABS stabilization. Wait until Section 5.5 to
program the Fluoro kV/mA tracking curves.
Содержание EMD Epsilon EPS 45-80
Страница 1: ...EPS 45 80 High Voltage Generator Technical Manual MANUAL P N 9894 00 G6...
Страница 81: ...3 Interfacing Technical Manual P N 9894 00 G6 3 23 Figure 3 6 2 2 Interlock X RAY shown in a RAD Fluoro...
Страница 142: ...EPS 45 80 High Voltage Generator 4 44 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...
Страница 244: ...EPS 45 80 High Voltage Generator 7 64 Technical Manual P N 9894 00 G6 THIS PAGE LEFT BLANK...