Section 5
Installation
5.1 Basic System Requirements
Power and Signal Requirements
Mains Connection:
115V AC, 60Hz or 230V AC, 50Hz (
10%)
Mains Wire Specification:
Number of Cores = 3
Current Rating minimum = 10 Amps
CSA (Cross Sectional Area minimum) = 1.50mm
2
Signal Wire Specification:
Number of Cores = 6 (+2 cores per additional signal)
Current Rating minimum = 1 Amp
CSA (Cross Sectional Area minimum) = 0.22mm
2
Power Consumption:
Maximum 100 W (VA)
Air Supply and Reagent Requirements
Instrument Air Requirements
Air Quality:
- 20°C dew point (free of water, oil and dust)
To meet or exceed the air quality specification, filter pack may be
required.
Air Supply Pressure:
1.5 bar
Air Supply Flow Rate:
Minimum 8.4 m
3
/hour at 1.5 bar
Air Consumption:
Average consumption is less than 5.4 m
3
/hour, typically 3.6 m
3
/hour
Typical Reagent Requirements
Acid Reagent: 6.0 N Sulfuric Acid (H
2
SO
4
), containing 350 mg/l Manganese Sulphate Monohydrate.
Base Reagent: 4.0 N Sodium Hydroxide (NaOH).
Sample, Drain and Exhaust Requirements
Sample Inlet & Outlet Pressure:
Ambient
Sample Inlet Temperature:
2°C
– 60°C (36°F - 140°F)
Sample Flow Rate:
Minimum 100ml per sample
Sample Particle Size:
Up to 100 microns
Drain & Exhaust:
Ambient
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...