Page 72
Preparation of Calibration Standard Solutions:
Use eye protection and gloves.
Standards solutions greater than 1000mg/l can be prepared directly without any dilution by simply mixing the
necessary quantity solvent or salt with deionized water. Standard solutions below 1000mg/l concentration
should be prepared by dilution technique. First a 1000mg/l standard stock solution should be prepared, and
then the required lower concentration standard solution should be prepared by applying the necessary dilution
procedures:
For example, to prepare a 50mgC/l TOC only standard solution, first weigh 50 grams of the
1000mgC/l stock standard. Add 50 grams of the 1000mgC/l standard into a one-liter flask and add
enough deionized water to make the solution exactly 1 liter.
For increased accuracy, standard solutions below 5mg/l (ppm) concentration should be prepared with
two or more steps dilution. For example, to prepare a 1mgC/l standard, first prepare a 100mgC/l
standard by adding 100 grams of the 1000mgC/l standard into a one-liter flask and by adding enough
deionized water to make it exactly a 1 liter solution. Then add 10 grams of the 100mgC/l standard into
a one-liter flask and add enough deionized water to make it exactly 1 liter.
Standard solutions at μg/l (ppb) levels should be prepared with several dilution steps. For instance, a
1mgC/l (1000 μg/l) standard should be prepared with two or more steps dilution as described above.
To prepare a 50μg/l standard, add 50 grams of the 1000μg/l standard into a one-liter flask, and add
enough deionized water to make it exactly 1 liter.
Shelf Life and Storage of Calibration Standard Solutions:
TOC standards prepared from Potassium Hydrogen Phthalate is typically stable for a month once it
is kept in a closed glass container and refrigerated at 4°C.
All other standards such as TOC prepared with Acetic Acid and TIC standards are recommended to
be used within 48 hours of manufacture.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...