8.1.3 Signal Simulate
S I G N A L S I M U L A T E 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
1 < * C H A N N E L 1 - 4 4 . 0 m A
5 A L A R M 1 - 3 O F F
8 C O 2 A L A R M 1 - 3 O F F
1 1 S Y N C R E L A Y O F F
1 2 S A M P L E S T A T U S 1 - 3 O F F
1 5 S A M P L E F A U L T 1 - 3 O F F
1 8 C A L S I G N A L O F F
1 9 M A I N T S I G N A L O F F
2 0 R E M O T E S T A N D B Y O F F
2 1 S T O P O F F
▼
S I G N A L S I M U L A T E 0 9 : 1 7 : 2 8 1 2 - 0 9 - 0 2
2 2 N O T E O F F
2 3 W A R N I N G O F F
2 4 F A U L T O F F
2 5 M A N M O D E T R I G O F F
2 6 4 - 2 0 m A C H N G O F F
2 7 4 - 2 0 m A C H N G 1 - 3 O F F
3 0 4 - 2 0 m A R E A D O F F
3 1
3 2 - - > I / O S T A T U S
This menu enables the user to test the Fault relay, the available 4-20mA outputs, the programmed output
signals and if installed, the Stream Alarm relays and any other optional outputs in the BioTector. Any settings
made by the user in this menu are automatically reset when the user exits this menu.
1. - 4. Channel 1-4.
This allows the user to test the function of each 4-20mA channel. Press the ENTER key,
use the arrow keys to set the required 4-20mA signal, and press the ENTER key again to test the 4-20mA
signal.
5. - 7. Alarm 1-3
(if configured in the system).
This allows the user to test the function of the stream
specific alarms if they are programmed in Output Devices menu. See Section
details. To change the state of the device, press the ENTER key, set the device to ON/OFF, and press
the ENTER key again. If the device is on, it will be marked with an “*”.
8. - 10. CO2 Alarm 1-3
(if configured in the system).
This allows the user to test the function of the stream
specific CO2 Alarms if they are programmed in Output Devices menu. See Section
for details. To change the state of the device, press the ENTER key, set the device to ON/OFF,
and press the ENTER key again. If the device is on, it will be marked with an “*”.
11. Sync Relay (if configured in the system).
Use this function to test the Synchronization Relay. To
change the state of the device, press the ENTER key, set the device to ON/OFF, and press the ENTER
key again. If the device is on, it will be marked
with an “*”. See Section
12. - 14. Sample Status 1-3 (if configured in the system).
Use this function to test the Sample Status
output signal for each specific stream. To change the state of the device, press the ENTER key, set the
device to ON/OFF, and press the ENTER key again. If the device is on, it will be marked with an “*”. See
Section
15. - 17. Sample Fault 1-3 (if configured in the system).
Use this function to test the Sample Fault output
signal for each specific stream. To change the state of the device, press the ENTER key, set the device
to ON/OFF, and press the ENTER key again. If the device i
s on, it will be marked with an “*”. See Section
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...