8.2 COMMISSIONING MENU
The Commissioning menus are used during the commissioning and startup of the analyzer. The functions in
the menus are used to program system site specific settings.
Commissioning Menu Diagram
8.2.1 Reaction Time
REACTION TIME 4m45s
REACTION TIME displays the total reaction time (in minutes and
seconds) for range 1, based on all programmed settings in
System Program, System Program 1 menu.
INTERVAL 0m
INTERVAL is the time (0 minutes by default) added between
each reaction. Interval can be programmed on site if frequent
analysis is not necessary. The longest interval time, which can
be programmed, is 1440 minutes (or 1 day). A programmed
interval time would reduce the reagent usage significantly.
When BioTector automatically extends the reaction time due to
high level of TIC and/or TOC in the sample, if any INTERVAL
time is programmed, the extended reaction time is taken out
from the interval time.
BioTector automatically generates an INTERVAL time when the
user programs any SAMPLER, FORWARD and/or REVERSE
times, which exceeds the maximum allowable time in the
Sample Pump menu below. The system determines the
maximum allowable time from the System Program 1 settings in
System Program menu.
TOTAL 4m45s
TOTAL displays the total reaction time including the interval time
if programmed.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...