5.2.4 System Fuse Specifications
Table 5 below summarizes the location and specification of the fuses used in BioTector. The locations of the
fuses are also displayed in figure 9 above.
Table 5 System Fuse Specifications
230 V
Systems
115 V
Systems
Location
Board
Number
Interru
pt
Rating
Type
Material
Fuse
Number
Current
Rating
Current
Rating
Power and
Input/Output
Board
81204350-01
H-250V
Miniature
5x20mm
Ceramic
F1
T 2.00A
T 3.15A
F1
*1
T 3.15A
T 5.00A
F2
T 1.00A
T 1.00A
F3, F4
T 500mA
T 500mA
F3, F4
*2
T 1.00A
T 1.00A
Motherboard 81204340_02
L-250V
Miniature
5x20mm
Glass
F2
F 2.00A
F 2.00A
F3
F 2.00A
F 2.00A
F4
F 2.50A
F 2.50A
F5
F 3.15A
F 3.15A
F6
F 2.50A
F 2.50A
BioTector
Compressor
Start Signal
Cable
N/A
L-250V
Miniature
5x20mm
Glass
N/A
T 200mA
T 200mA
KEY
A:
Amperes
DIN:
German Institute for Standardization (Deutsches Institut für Normung e.V.)
F1-6:
Fuse Number
F:
Fast Acting (Fast Blow)
H:
High Interrupt
ID:
Identification
L:
Low Interrupt
mA:
Milli-amperes
N/A:
Not Applicable
PCB:
Printed Circuit Board
T:
Time Lag (Time Delay)
V:
Volts
*1:
(Optional) used with TDK power supply, type SWS300-24 only.
*2:
(Optional) if higher powered device are connected, for example Multi-channel Modbus / Profibus,
External Valve, External Relay.
BioTector contains electrical components operating under high voltages.
Contact may result in electric shock and severe or fatal injury.
All electrical work should be carried out by qualified electrical personnel only.
When any fuse replacement is required in the system, please refer to table 5
below and system specific drawings.
DANGER
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...