Page 14
1.5.4 Sample Stream Precautions
The user is responsible to establish the potential hazard associated with each sample stream. Necessary
precautions must be taken, to avoid physical contact with any harmful sample stream, which may contain
chemical or biological hazards.
System components and their composition, which come in contact with the sample liquid and possible volatile
gases from the sample, are tabulated in table 1 below. If there are suspected compatibility issues between the
sample stream and BioTector components, please contact the manufacturer or the distributor.
Table 1 System components and their composition
Component
Material
Tubing
PFA (Per-fluoro-alkoxy)
Vinyl
Reinforced Silicon
Reinforced PVC (Poly-vinyl-chloride)
Fittings
PFA (Per-fluoro-alkoxy)
Stainless Steel (SS-316)
PVDF (Poly-vinylidene-flouride)
Brass
Pump Tubing
EMPP (Elastomer-modified-poly-propylene)
Connectors
PP (Poly-propylene)
HDPE (High-density-poly-ethylene)
Connector & Valve Tubing
EMPP (Elastomer-modified poly-propylene)
Viton / FKM (Fluoro-elastomer)
Sample Valve
PEEK (Poly-ether-ether-ketone)
PVDF (Poly-vinylidene-flouride)
Stainless Steel (SS-316)
EMPP (Elastomer-modified poly-propylene)
Reactor
Hastelloy (C-276)
Stainless Steel (SS-316)
PFA (Per-fluoro-alkoxy)
PTFE (Poly-tetra-fluoro-ethylene)
Valve Seals
Kalrez / FFKM (Perfluoro-elastomer)
Viton / FKM (Fluoro-elastomer)
NBR (Nitrile-butadiene-rubber)
Oxidized Sample Catch-pot/Cleaning Vessel
Borosilicate Glass
NDIR CO
2
Analyzer
Hastelloy (C-276)
Stainless Steel (SS-316)
NDIR CO
2
Analyzer Lens
Sapphire
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...