Page 35
CONSUMABLES
Typical Replacement Frequency & Consumption
Acid & Base:
26 weeks/19 Liters (based on standard range and 6 minute analysis cycle)
Instrument Air:
1.5 bar, - 20°C dew point (free of water, oil and dust)
Average consumption is less than 5.4 m
3
/hour.
Filter pack is recommended and available to meet the air quality specification.
BioTector Air Compressor is available for air supply.
Service:
6 Monthly Intervals
ANALYSIS PARAMETERS
Oxidation Method:
Patented Two-Stage Advanced Oxidation Process using Hydroxyl Radicals
TOC Measurement:
NDIR measurement of CO
2
after oxidation
Measurement Terms:
TOC (Total Organic Carbon) including Non-Purgeable Organic Carbon
(NPOC) and Purgeable Organic Carbon (POC)
BioTector TIC&TOC mode measures NPOC
BioTector VOC mode measures TOC as NPOC+POC.
Measured Components:
TOC (NPOC)
TOC (NPOC + POC)
TIC
TC
VOC (POC)
COD*
BOD*
*
COD & BOD by correlation algorithm incorporating TOC measured results
Cycle Time:
TOC typically 6 minutes
MONITORING RANGES:
TOC
Standard Range
0-5mgC/l
or
0-25mgC/l
Accuracy and Repeatability
3% of reading or
0.03mgC/l, whichever is greater
Detection Limit
0.06mg/l
Second Range
(Optional for Exceedance Tracking)
0-100mgC/l (
5% of reading or
0.5mgC/l, whichever is greater
)
Range Selection:
Automatic or Manual (up to 2 ranges configurable)
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...