52_HIGH CO2 IN BASE
CO
2
level in the base reagent, which is
checked during the Zero Calibration or
Zero Check cycles, is higher than the
“BASE CO2 ALARM” level
programmed in Fault Setup menu.
Confirm the base reagent CO
2
filter is serviced and in good
condition, and the container is
sealed.
Check reagent quality.
Replace the base reagent.
62_SMPL PUMP STOP ON
Sample Pump stopped with its rotation
sensor ON or the sensor is faulty and it
is always showing ON.
Run the Sample Pump and check
the rotation.
Check pump sensor signal looking
at DI15 in Digital Input menu.
Replace the pump.
63_SMPL PUMP STOP OFF
Sample Pump stopped with its rotation
sensor OFF or the sensor is faulty and
it is not sensing the rotation of the
pump.
Run the Sample Pump and check
the rotation.
Check pump sensor signal looking
at DI15 in Digital Input menu.
Replace the pump.
81_ATM PRESSURE HIGH
System atmospheric pressure sensor
reading is above 115kPa. Therefore,
the sensor reading is defaulted to
101.3kPa fault operation mode.
Check ADC[05] in Analog Input
menu, which should be ~4 Volts.
Faulty pressure sensor.
Replace the motherboard.
82_ATM PRESSURE LOW
System atmospheric pressure sensor
reading is below 75kPa. Therefore, the
sensor reading is defaulted to
101.3kPa fault operation mode.
Check ADC[05] in Analog Input
menu, which should be ~4 Volts.
Faulty pressure sensor.
Replace the motherboard.
83_SERVICE TIME
The service counter has counted down
the days between service intervals and
indicates that service is required.
Service the BioTector.
After service, acknowledge the
warning by resetting the counter
using the “RESET SERVICE
COUNTER” function in
Diagnostics, Service menu.
84_SAMPLER ERROR
Warning generated on the BioTector
venturi/vacuum sampler due to no/low
sample or low air pressure/vacuum in
the sampler.
Check the LCD screen of the
venturi/vacuum sampler for
details.
See BioTector venturi/vacuum
sampler user manual.
88_O2 CONTROLLER
WARN
A problem has been found during
communication with the O2-Controller
Board.
The communication between the O2-
Controller Board and the Motherboard
is lost.
Check LED 2 (L2) on the O2-
Controller Board, it should be on.
Check the 24V DC power on the
O2-Controller Board at terminals
N01 and N02.
Check the data cable (ribbon
cable) connections on the board.
Power down and power up the
BioTector to reset the system.
Replace the board.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...