Page 139
12_HIGH CO2 IN O2
High level of CO
2
has been detected in
the input oxygen gas.
Go to Simulate menu and read the CO
2
analyzer
’s CO
2
ppm value. If this value
is greater than 250-300ppm, check the
oxygen purity.
Check the quality of oxygen
following the Oxygen Purity Test
procedures in
If the oxygen purity test is
satisfactory, open CO
2
analyzer
and clean the optics.
If problem persists, replace the
CO
2
Analyzer filters.
If the oxygen purity test is
unsatisfactory, replace the oxygen
concentrator.
18_LIQUID LEAK DET
BioTector’s liquid leak detector is
activated by the three liquid leak
detector points fitted in the BioTector.
First point is in the main analysis
enclosure cabinet, the second point is
in the oxygen concentrator cabinet and
the third point is in the Mixer Reactor.
See figure 2 in Section
Check for a liquid leak in the
BioTector main cabinet and in the
oxygen concentrator cabinet.
Check for a liquid leak in Mixer
Reactor by unplugging the leak
detector connector on the bottom
of the reactor.
Rectify the leak.
20_NO REAGENTS
(can also be programmed as a
warning or notification condition)
BioTector has calculated that the
reagent containers should be empty.
Change reagents, and reset
reagent monitor in Install New
Reagents menu.
Confirm that the reagent
volume/container size input values
are correct in the Reagents
Monitor menu.
If necessary, to readjust the
number of days reagents last, see
Section
for details.
104_MOTHERBOARD
FUSE4
Fuse 4
(for Cooler and Ozone Generator)
is blown on the motherboard. See
Section
Check/replace fuse 4 (F4).
Check fuse holder and confirm
that the fuse is snapped in tightly
into the fuse holder.
105_MOTHERBOARD
FUSE5
Fuse 5
(for sample/acid/base pumps, mixer
reactor, sample/stream/manual valves,
reactor/exhaust valves, cooler fan, ozone
destructor, ozone generator fan)
is blown
on the motherboard. See Section
details.
Check/replace fuse 5 (F5).
Check fuse holder and confirm
that the fuse is snapped in tightly
into the fuse holder.
Go to Simulate menu and activate
each device connected.
106_MOTHERBOARD
FUSE6
Fuse 6
(which is reserved for future)
is
blown on the motherboard. See Section
details.
Check/replace fuse 5 (F5).
Check fuse holder and confirm
that the fuse is snapped in tightly
into the fuse holder.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...