Page 138
04_REACTION ERROR
(can also be programmed as a
warning condition)
No TOC (or TC) CO
2
peak detected or
the CO
2
peak is below the “CO2
LEVEL” for 3 consecutive reactions.
See Section
Acid/Base containers are empty.
Acid/Base tube rail installed
incorrectly.
Acid/Base Pumps are faulty.
Delivery problem or air bubbles in
Acid/Base lines.
Mixer Reactor is not mixing.
Acid/Base reagents are at
incorrect strength.
Run pH Test (see Section
05_PRESSURE TEST FAIL
MFC flow did not drop below the
“PRESSURE TEST FAULT” level
during the pressure test cycle. See
Section
Gas/liquid leak in the BioTector.
Valve leaking or not sealing.
Open Reactor Valve, inspect for
dirt/damage.
Check system fittings.
Check Mixer Reactor.
Run Pressure Test (see Section
details).
06_PRESSURE CHCK FAIL
MFC flow did not drop below the
“PRESSURE CHCK FAULT” level
during the pressure check cycle for 3
consecutive reactions. See Section
8.3.4.5 Pressure/Flow Test Program
for details.
Gas/liquid leak in the BioTector.
Valve leaking or not sealing.
Open Reactor Valve, inspect for
dirt/damage.
Check system fittings.
Check Mixer Reactor.
Run Pressure Test (see Section
details).
11_CO2 ANALYZER FAULT
Fault in CO
2
Analyzer.
Very dirty optics in CO
2
Analyzer.
Check CO
2
analyzer
’s ppm CO
2
response in Simulate menu (see
Section
details).
Open CO
2
analyzer and clean the
optics.
Power down and power up the
BioTector.
If problem persists, check
outgoing 24V DC power to CO
2
analyzer on Motherboard at wires
N11 and N12.
For further tests, see information
sheet “T019. BioTector CO
2
Analyzer
Troubleshooting”, which is available in
the MMC/SD card shipped with the
BioTector.
Содержание BioTector B3500C
Страница 17: ...Page 17 Software Menu Diagram...
Страница 44: ...Page 44 Figure 4 BioTector analysis layout typical TIC TOC system...
Страница 46: ...Page 46 Figure 6 BioTector oxygen concentrator layout...
Страница 63: ...Page 63...
Страница 78: ...Page 78 Section 8 Maintenance Menu Maintenance Menu Diagram...
Страница 155: ...Page 155 Section 11 System Replacement and Spare Parts...
Страница 163: ...Page 163 ZK Zero check ZM Manually input zero adjust ZS Zero and Span...