P A R T F O U R - S E R V I C E A N D M A I N T E N A N C E
S
ECTION
1
1.1 Inspecting Sensor
Cable
1.2 Removing/Attaching
Analyzer Door
If a measurement problem exists and you suspect the sen-
sor cable, inspect it for physical damage. If an interconnect
cable is used, disconnect the cable at both ends (junction
box and analyzer) and, using an ohmmeter, check its wires
for continuity and internal shorts.
To Remove Analyzer Door:
1. Loosen the four screws on the analyzer’s front bezel
and swing open the door.
2. Disconnect the ribbon cable connector from the back-
side of the door.
3. While holding the door, slide the door release lever
(Figure 4-1) upward to release the door from the hinge.
To Attach Analyzer Door (Figure 4-1):
1. With the door open, hold the spring-loaded door re-
lease lever upward, carefully position the hinge slots
onto the hinge until the hinge is fully seated, and re-
lease the lever.
2. After the door is attached, connect the ribbon cable
connector to the backside of the door. To prevent acci-
dentally crimping the cable, push any excess cable into
the opening by the connector.
3. Close the door and tighten the four screws on the ana-
lyzer’s front bezel.
FIGURE 4-1 Analyzer Door Attachment Details