PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
73
CAL BUFFER VALUES
pH STABILITY MODE
Use the following “PULSE SUPPRESSION” submenu to
turn this feature on or off:
Only use the following “CAL BUFFER VALUES” submenu
when you intend to use the “Pre-Defined Buffer Method” for
pH calibration:
Choose the specific buffer set you want to use for pH cali-
bration, and then select the desired low and high buffer
values from that set. For calibration details using the “Pre-
Defined Buffer Method,” refer to PART THREE, Section 4.4
under the respective subheading.
Use the following “pH STABILITY MODE” submenu to set
the acceptable rate of stability for the measured pH and
temperature signals during calibration:
•
LOW:
During calibration, the pH and
°
C annunciators
stop flashing when these two measured signals are less
stable (changing moderately).
•
MED:
During calibration, the pH and
°
C annunciators
stop flashing when these two measured signals are
more stable (less change per time period than “LOW”
setting).