PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
57
5.4 Establishing Wash/Cal
System Operation
☞
For GLI
Cal-Clean™ System
NOTE:
Depending on whether the analyzer is equipped for
use with a conventional combination electrode or a
GLI Differential Technique pH sensor, it has differ-
ent software to operate the wash system:
•
Combination Electrode:
The analyzer software
provides wash and calibration functions, specifi-
cally adapted for use with a GLI Cal-Clean™
System (or other appropriate hardware).
•
Differential Technique pH Sensors:
Different
analyzer software provides a wash only function,
specifically adapted for use with a GLI Air/Water
Blast Cleaning System.
Refer to the appropriate subsection for software
operating details.
The analyzer software provides wash and calibration func-
tions that require a GLI Cal-Clean™ System (or equivalent
engineered automatic system) to control appropriate hard-
ware. The wash/cal functions can be used two basic ways:
1. For Continuous Measurement with Periodic Wash/Cal,
enabling you to automatically:
•
Retract measuring electrode from the process
•
Dispense wash and/or rinse solutions to clean electrode
•
Dispense buffers for calibration
•
Calibrate (single or two-point calibration)
•
Insert cleaned and calibrated electrode into the process
2. For Continuous Retraction with Periodic Measurement,
enabling you to automatically:
•
Insert measuring electrode into the process
•
Obtain measurements for a defined time period
•
Retract the measuring electrode from the process
•
Dispense wash and/or rinse solutions to clean electrode
•
Dispense buffers for calibration
•
Calibrate (single or two-point calibration)
•
Store cleaned and calibrated electrode in the positioner