PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Model P63 pH Analyzer
Rev. 7-1201
58
The analyzer software enables you to create “Wash” and
“Cal” cycles containing specific function steps to meet ap-
plication requirements. The established cycles can then be
scheduled for specific times and days, or set to occur at re-
petitive intervals. Wash and Cal cycles can also be started
by a remote TTL input signal after enabling the remote
system control feature.
Dedicated Cal-Clean™ System
Outputs and Required Inputs
When the wash/cal system feature is enabled, normal relay
and TTL output operation is suspended and replaced by
dedicated wash/cal operation. All of the analyzer outputs
are used to control wash/cal functions:
Analyzer Output
Wash/Cal Control Function
Relay A ............................Rinse
Relay B ............................Auxiliary Wash 1 and 2
Relay C............................Main Wash
TTL Output A ...................Sensor Retract
TTL Output B ...................Buffer 1
TTL Output C...................Buffer 2
Also, both analyzer TTL inputs are dedicated to monitoring
the pH electrode position:
TTL Input
Input Signal State
X......................................Low when electrode is inserted
Y......................................Low when electrode is retracted
To enable the wash/cal system feature, press
CONFIG key
to display the CONFIGURE main menu, select the
“Wash/Cal System” line, press
ENTER key
to display the
following submenu screen, and select “YES.”
The “Define Wash Cycle” and “Define Cal Cycle” submenus
are used to respectively configure “Wash” cycles and “Cal”
cycles. Both submenus contain many of the same function
steps but also have different steps for their specific pur-
poses (see respective subsection for details). Although
many wash and cal function steps have the same names
and functions, they operate independently of each other
and must be separately configured.