PART THREE - OPERATION
SECTION 6 - MAINTENANCE MENU
Model P63 pH Analyzer
Rev. 7-1201
80
6.2 HOLD/XFER OUTPUTS
☞
6.3 MANUAL PID
(manually operate
PID controller)
After selecting “Hold/Xfer Outputs” from the MAINTENANCE
menu, use the following submenu to manually change the
output modes:
When cleaning the sensor or performing other maintenance
tasks, you may want to hold the analog outputs and relays
at their present states or transfer them to states that will not
adversely affect the connected devices. By selecting
“HOLD” or “XFER,” the output states respond accordingly
even after exiting this submenu screen. The HOLD or XFER
annunciator will appear on the MEASURE screen to remind
you that this condition exists. To cancel the manual hold or
transfer, use this same submenu screen and then select
“ACTIVE.”
NOTE:
Using this screen to manually hold or transfer re-
lays may interact with relay overfeed timer
operation (if counting). Refer to PART THREE,
Section 8.5 for complete details.
After selecting “Manual PID” from the MAINTENANCE
menu, use this submenu to manually operate the PID con-
troller:
By manually operating the PID controller, you can check the
control elements for proper sizing, and get a “feel” for the
capability of the control system.