PART THREE - OPERATION
SECTION 5 - CONFIGURE MENU
Rev. 7-1201
Model P63 pH Analyzer
63
☞
•
Buffer 2:
Sets how long (0-999 seconds) the “Buffer 2”
step is to remain on. If “0” is entered, this second calibra-
tion point step (and its related “Buffer 2 Delay” step, even
if it has an entered time) will be bypassed in the cal cycle.
•
Buffer 2 Delay:
Sets the time period (0-999 seconds)
between the “Buffer 2” step and calibration of the second
point. This entered time can be used to allow the pH
reading to stabilize before attempting to calibrate this
second point. If a “0” is entered, this step will be bypassed
and the “Cal” cycle will proceed with the third “Rinse” and
“Rinse Delay” steps.
NOTE:
If only a single-point calibration is desired, enter a
time of “0” seconds for the “Buffer 1” or “Buffer 2”
steps to bypass that respective calibration point.
After Buffer 2 is delivered, the analyzer automatically ini-
tiates calibration of the second point. After the second
point has been successfully calibrated, the third “Rinse”
and “Rinse Delay” steps will occur for their respectively
entered times before the “Cal” cycle continues with the
“Aux Wash 2” step.
•
Aux Wash 2:
See “Wash” cycle section for description.
•
Aux Wash 2 Delay:
See “Wash” cycle section for description.
SCHEDULE WASH/CAL Events
for GLI Cal-Clean™ System
Use the “SCHEDULE WASH/CAL” submenu screen (exam-
ple screen shown below) to schedule when you want
“Wash” and “Cal” cycles to occur:
Up to 28 separate events (“Wash” or “Cal” cycles) can be
scheduled. Each line on the “SCHEDULE WASH/CAL”
submenu represents one event. To schedule an event, en-
ter three items:
•
Day of week:
Selections are “SUN,” “MON,” “TUE,”
“WED,” “THU,” “FRI,” “SAT,” or “EVR” for an event re-
curring every day.