PART ONE - INTRODUCTION
SECTION 2 - SPECIFICATIONS
Model P63 pH Analyzer
Rev. 7-1201
16
S
ECTION
2
2.1 Operational
Display....................................... Graphic dot matrix LCD, 128 x 64 pixels with
LED backlighting; 1/2 inch (13 mm) main dis-
play character height; 1/8 inch (3 mm)
auxiliary information character height; menu
screens contain up to six full lines of text
Displayed Information
Ranges
Main Display........................... 0.0-14.0 pH or 0.00-14.00 pH, selectable
Auxiliary Display:
Temperature........................ -10.0 to +110.0
°
C
Opt. PID Controller Output .. 0.0-100.0%
Sensor Status...................... Displays predicted date for end-of-sensor life
Date .................................... Month/day/year
Time.................................... Hour/minutes
Calibration Status................ Displays next scheduled date for calibration
mA Outputs (1 and 2) .......... 4.00-20.00 mA
Act. Electrode Impedance.... 1-999 M
Ω
(Differential Technique sensor)
Std. Electrode Impedance.... 1-999 M
Ω
(Differential Technique sensor)
Glass Elect. Impedance....... 1-999 M
Ω
(conventional combination electrode)
Ref. Electrode Impedance.... 1-250 K
Ω
(conventional combination electrode)
pH Sensor mV Output ......... -500 to +500 mV
Relay Status........................ On or off
Diagnostic Warnings ........... See Table A for details
Error Messages ................... See Table A for details
Ambient Conditions .................... -22 to +140
°
F (-30 to +60
°
C); 0-95% relative
humidity, non-condensing
Relays:
Types/Outputs: Standard ....... Three electromechanical relays (two SPDT
and one SPST); UL-rated 5A 115/230 VAC,
5A @ 30 VDC resistive
Optional ..........Three solid state AC relays (all SPST); UL-rated
2A continuous; user must provide 24-250 VAC
and 0.02 amps RMS minimum
or
Three solid state DC relays (all SPST); UL-rated
2A continuous; user must provide 3-60 VDC
Functional Modes ................... Each relay (A, B, and C) can be selected to be
driven by the measured pH or temperature;
or all relays operate as dedicated wash/cal
cycle outputs for Cal-Clean™ System
Operating Modes: Control....... Settings for fail safe on/off, high/low phasing,
setpoint, deadband, overfeed timer, on de-
lay, and off delay
Alarm...........Settings for fail safe on/off, high alarm point, high
alarm point deadband, low alarm point, low
alarm point deadband, on delay, and off delay
Indicators................................ Relay A, B, and C annunciators indicate
respective relay on/off status
Temperature Compensation ....... Automatic or manual, -10.0 to +110.0
°
C, with
selection for temperature element (NTC 300
ohm thermistor, Pt 1000 ohm RTD or Pt 100
ohm RTD), or a manually entered value; 3-
wire temp. sensor connection capability; auto
temp. comp. for pure water (0.0-50.0
°
C)
SPECIFICATIONS